American Journal of Clinical Nutrition, Vol 63, 157-163, Copyright © 1996 by The American Society for Clinical Nutrition, Inc
ORIGINAL RESEARCH COMMUNICATIONS |
JC Picaud, J Rigo, K Nyamugabo, J Milet and J Senterre
Department of Neonatology, University of Liege, Belgium.
The reproducibility, accuracy, and precision of dual-energy X-ray absorptiometry (DXA) was assessed by scanning 13 piglets (1471-5507 g) in triplicate. In four piglets, fat content was increased with porcine lard around the abdomen; additional measurements were performed on these animals. Reproducibility in DXA measurements from the animals without added fat was 0.09% for body weight, 1.95% for bone mineral content (BMC), and 5.35% for fat content. DXA estimates of body weight, BMC, and fat content were significantly correlated with scale body weight, ash weight, chemical calcium, and chemical fat. Body weight was measured accurately but fat content was overestimated by DXA. Mean BMC estimated by DXA represented 48% of ash weight and 215% of calcium content. The precision of DXA was 0.23% for body weight, 10.99% for ash weight, and 4.44% for calcium content. The precision of DXA for fat content was poor. However, for measurements performed in piglets with > 250 g fat, the precision was 8.85%. Thirty appropriate-forgestational- age term human neonates (birth weight: 3188 +/- 217 g) were scanned once during the first week of life. BMC and fat content were 54 +/- 6 and 470 +/- 92 g, respectively, which corresponded to 26.4 +/- 2.6 g calcium and 427 +/- 82 g fat. These were close to the reference values previously determined by chemical analysis. This study suggests that DXA is accurate and reliable for measurement of calcium and fat contents in human neonates. Further refinements would be beneficial for determining fat content in preterm human infants.
This article has been cited by other articles:
![]() |
K. J Ellis, M. Yao, R. J Shypailo, A. Urlando, W. W Wong, and W. C Heird Body-composition assessment in infancy: air-displacement plethysmography compared with a reference 4-compartment model Am. J. Clinical Nutrition, January 1, 2007; 85(1): 90 - 95. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. W. Koo and E. M Hockman Posthospital discharge feeding for preterm infants: effects of standard compared with enriched milk formula on growth, bone mass, and body composition Am. J. Clinical Nutrition, December 1, 2006; 84(6): 1357 - 1364. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Koo Maternal calcium supplementation and bone accretion in infants. Am. J. Clinical Nutrition, October 1, 2006; 84(4): 943 - 943. [Full Text] [PDF] |
||||
![]() |
M. T. Clandinin, B. Larsen, J. V. Aerde, and W. W. K. Koo Reduced Bone Mineralization in Infants Fed Palm Olein-Containing Formula: A Randomized, Double-Blinded, Prospective Trial Pediatrics, September 1, 2004; 114(3): 899 - 900. [Full Text] [PDF] |
||||
![]() |
W. W. K. Koo, E. M. Hockman, and M. Hammami Dual Energy X-Ray Absorptiometry Measurements in Small Subjects: Conditions Affecting Clinical Measurements J. Am. Coll. Nutr., June 1, 2004; 23(3): 212 - 219. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Ma, M. Yao, Y. Liu, A. Lin, H. Zou, A. Urlando, W. W Wong, L. Nommsen-Rivers, and K. G Dewey Validation of a new pediatric air-displacement plethysmograph for assessing body composition in infants Am. J. Clinical Nutrition, April 1, 2004; 79(4): 653 - 660. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. W. K. Koo, M. Hammami, and E. M. Hockman Validation of Bone Mass and Body Composition Measurements in Small Subjects with Pencil Beam Dual Energy X-Ray Absorptiometry J. Am. Coll. Nutr., February 1, 2004; 23(1): 79 - 84. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Chauhan, W. W. K. Koo, M. Hammami, and E. M. Hockman Fan Beam Dual Energy X-Ray Absorptiometry Body Composition Measurements in Piglets J. Am. Coll. Nutr., October 1, 2003; 22(5): 408 - 414. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. W. Koo, M. Hammami, and E. M Hockman Interchangeability of pencil-beam and fan-beam dual-energy X-ray absorptiometry measurements in piglets and infants Am. J. Clinical Nutrition, August 1, 2003; 78(2): 236 - 240. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D Sainz and A. Urlando Evaluation of a new pediatric air-displacement plethysmograph for body-composition assessment by means of chemical analysis of bovine tissue phantoms Am. J. Clinical Nutrition, February 1, 2003; 77(2): 364 - 370. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. R. Schmelzle and C. Fusch Body fat in neonates and young infants: validation of skinfold thickness versus dual-energy X-ray absorptiometry Am. J. Clinical Nutrition, November 1, 2002; 76(5): 1096 - 1100. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hammami, J.-C. Picaud, C. Fusch, E. M. Hockman, J. Rigo, and W. W.K. Koo Phantoms for Cross-Calibration of Dual Energy X-ray Absorptiometry Measurements in Infants J. Am. Coll. Nutr., August 1, 2002; 21(4): 328 - 332. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Koo, J. Walters, E. Hockman, and W. Koo Body Composition of Newborn Twins: Intrapair Differences J. Am. Coll. Nutr., June 1, 2002; 21(3): 245 - 249. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. W. K. Koo, M. Hammami, and E. M. Hockman Use of Fan Beam Dual Energy X-Ray Absorptiometry to Measure Body Composition of Piglets J. Nutr., June 1, 2002; 132(6): 1380 - 1383. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L Salle, E. E Delvin, A. Lapillonne, N. J Bishop, and F. H Glorieux Perinatal metabolism of vitamin D Am. J. Clinical Nutrition, May 1, 2000; 71(5): 1317S - 1324. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. C. Mehta, B. L. Specker, S. Bartholmey, J. Giddens, and M. L. Ho Trial on Timing of Introduction to Solids and Food Type on Infant Growth Pediatrics, September 1, 1998; 102(3): 569 - 573. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. W. K. Koo, A. J. Bush, J. Walters, and S. E. Carlson Postnatal Development of Bone Mineral Status During Infancy J. Am. Coll. Nutr., February 1, 1998; 17(1): 65 - 70. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. S. Rose, W. P. Flatt, R. J. Martin, and R. D. Lewis Whole Body Composition of Rats Determined by Dual Energy X-Ray Absorptiometry Is Correlated with Chemical Analysis J. Nutr., February 1, 1998; 128(2): 246 - 250. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |