Original Research Communications |
Background: Previous research has suggested that there may be significant within-subject variability, both site-to-site and over time, in hemoglobin concentrations in capillary blood.
Objective: This study examined the reliability of the portable hemoglobinometer (PHM) system with use of capillary blood and the implications of errors of the magnitude found for the classification of anemia status in individuals and population groups. The precision and accuracy of the method with use of venous blood were also tested.
Design: Three empirical data sets were used to measure reliability, precision, and accuracy of the PHM system [2 from Honduras (n = 87 and 141); 1 from Bangladesh (n = 73)]. Simulation data were used to assess the implications of errors for screening individuals for anemia and to estimate anemia prevalence.
Results: High within-subject variability (unreliability) was identified when capillary blood from the left hand was compared with that from the right hand (CV: 6.3%) and when measurements were taken on 4 consecutive days (CV: 7.0%). Reliability was only 69% and 50%, respectively. Precision and accuracy, however, were very high (concordance coefficients of 0.99 and 0.98 and CV < 1%).
Conclusions: The simulation data showed that errors of the magnitude found due to unreliability can lead to misclassification of anemia status in individuals and small biases in anemia prevalence estimates. We recommend replicate sampling to reduce the influence of unreliability in the use of the PHM system with capillary blood.
Key Words: Nutrition assessment hemoglobin error reliability precision accuracy capillary blood hemoglobinometer Honduras Bangladesh anemia
This article has been cited by other articles:
![]() |
T. W. McDade and B. Shell-Duncan Whole Blood Collected on Filter Paper Provides a Minimally Invasive Method for Assessing Human Transferrin Receptor Level J. Nutr., December 1, 2002; 132(12): 3760 - 3763. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. W Solomons and K. Schumann Lessons learned in iron intervention trials Am. J. Clinical Nutrition, September 1, 2002; 76(3): 691 - 693. [Full Text] [PDF] |
||||
![]() |
A. Bhargava, H. E. Bouis, and N. S. Scrimshaw Dietary Intakes and Socioeconomic Factors Are Associated with the Hemoglobin Concentration of Bangladeshi Women J. Nutr., March 1, 2001; 131(3): 758 - 764. [Abstract] [Full Text] |
||||
![]() |
P. R Pehrsson, P. B Moser-Veillon, L. S Sims, C. W Suitor, and E. Russek-Cohen Postpartum iron status in nonlactating participants and nonparticipants in the Special Supplemental Nutrition Program for Women, Infants, and Children Am. J. Clinical Nutrition, January 1, 2001; 73(1): 86 - 92. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Zavaleta, L. E Caulfield, and T. Garcia Changes in iron status during pregnancy in Peruvian women receiving prenatal iron and folic acid supplements with or without zinc Am. J. Clinical Nutrition, April 1, 2000; 71(4): 956 - 961. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |