| ABSTRACT |
|---|
|
|
|---|
Objective: Our objective was to determine the bias and agreement between DXA and a 4-compartment model in predicting the percentage of fat mass (%FM) in a multiethnic group of young females.
Design: The %FM values measured by DXA of 73 white, 43 African American, 14 Hispanic, and 11 Asian females with a mean (± SD) age of 13.0 ± 1.9 y were compared with the 4-compartment values, which were based on measurements of body density, body water, and bone mineral content.
Results: The %FM values measured by the 2 methods were correlated at r = 0.90 with an SEE of 3.3%; Bland-Altman analysis indicated an average bias of 3.9%. After nullification of the average bias, an individual estimate of %FM by DXA could be underestimated or overestimated by 6.7% when compared with the 4-compartment value.
Conclusions: DXA is an appropriate method for estimating body composition in a group of young females because its bias and limits of agreement are independent of age, ethnicity, and body fatness. However, the limits of agreement of 6.7% could cause an individual FM value to be underestimated or overestimated by 28% relative to the 4-compartment value. Therefore, DXA may not be the optimal method of measuring the body fatness of young females.
Key Words: Body fat percentage of fat mass fat-free mass girls female adolescents densitometry isotope dilution dual-energy X-ray absorptiometry 4-compartment model
| INTRODUCTION |
|---|
|
|
|---|
Recently, on the basis of data collected on children and adolescents in the 1999 National Health and Nutrition Examination Survey (NHANES), officials from the National Center for Health Statistics (NCHS) voiced concern that the increase in obesity among Americas youth that had begun in the 1980s appears to be continuing (4). Because body weight in adulthood is strongly associated with body weight in adolescence (5), the NCHS interpreted the most recent NHANES data as suggesting the likelihood of another generation of adults who may be at risk for overweight and obesity-related health conditions. Because body weight and body mass index are poor indicators for assessing the true degree of adiposity (6), it is essential to determine the percentage of body fat with the most accurate methodology.
The aim of this study was to evaluate the level of agreement between DXA and the 4-compartment criterion model in estimating body fat in a multiethnic group of young females.
SUBJECTS AND METHODS
Human subjects
A total of 141 young females (73 whites, 43 African Americans, 14 Hispanics, and 11 Asians) participated in the study. The young females, who were between 9 and 17 y of age, were recruited from schools in the greater Houston metropolitan area. All subjects were healthy and nondiabetic at the time of the study. The Institutional Review Board for Human Research at Baylor College of Medicine approved the protocol. All subjects and their parents gave written, informed consent. The study was part of a larger study to define the body composition and energy metabolism of pregnant teenagers. None of the young females in the present study were pregnant.
Sexual maturity evaluation
Sexual maturity was determined from a physical examination by a physician according to the Tanner stages of classification (7).
Anthropometric measurements
On admission to the Childrens Nutrition Research Center, each subjects body weight was measured to the nearest 0.1 kg with an electronic scale (Scale-Tronix, Wheaton, IL), and height was measured to the nearest 0.1 cm with a stadiometer (Holtain Ltd, Croswell, Crymych, United Kingdom). One investigator (JES) made all the anthropometric measurements.
Dual-energy X-ray absorptiometry
Fat mass (FM), fat-free mass (FFM), and bone mineral content (BMC) were measured by DXA (Hologic QDR-2000W, software version 5.56; Hologic, Inc, Waltham, MA) with the pencil beam. The Hologic instrument uses a single scan mode for all subjects aged
8 y. The fundamental principles and operating procedures of DXA have been extensively described in the literature (1, 813) and will not be presented here.
Four-compartment criterion method
The reference values for percentage of fat mass (%FM) were obtained by using a 4-compartment model as follows (14)
![]() | (1) |
For TBW measurement, a baseline plasma sample was collected by venipuncture from all subjects before they drank 1.25 g 10% (by wt) H218O (Isotec Inc, Miamisburg, OH)/kg body wt. Another plasma sample was collected 3 h later. The plasma samples were prepared for oxygen isotope ratio measurements by gas-isotope-ratio mass spectrometry (18). TBW was calculated as follows
![]() | (2) |
Statistical analysis
Linear regression analysis and the Bland-Altman procedure (21) were used to compare %FM values obtained by DXA (%FMDXA) with those obtained by the 4-compartment method (%FM4C). In the regression analysis, %FMDXA was plotted against %FM4C. The deviation of the slope from 1.0 and the deviation of the intercept from zero of the regression line were evaluated by using the critical t values with the corresponding degrees of freedom. The error of the DXA method was defined by the SEE of the regression line. With the Bland-Altman procedure, differences between the 2 methods were plotted against the averages of the 2 methods. Regression analysis was used to examine the relation between the differences and averages. If the slope was not significant (P > 0.05), the bias (mean difference between methods) and the 95% limits of agreement [bias ± (2 x the SD of the differences)] were computed. All statistical analyses were performed with SPSS for WINDOWS (version 8; SPSS Inc, Chicago).
RESULTS
Mean values for age, sexual maturation, physical characteristics, and body composition of the 141 young females are shown in Table 1
. Because of the small number of young Hispanic and Asian females who participated in the study, Tanner-stage data are presented only for those with breast and pubic hair development at Tanner stage
3. Most of the young females were at Tanner stage
3. One-way analysis of variance and Tukeys test showed that the young African American females were significantly heavier and taller than the young white females (P < 0.05). The young African American females also had significantly higher body mass indexes, BMC, TBW, FFM, and BMC/FFM than did the young white females (P < 0.05). Within each ethnic group, the FFM, BMC/FFM, TBW/FFM, and FM estimated by DXA were significantly different from those estimated by the 4-compartment model (P < 0.05). On average, DXA overestimated %FM by 3.9 ± 3.4% (P < 0.01) relative to the 4-compartment model.
|
|
0.28). Univariate analysis of variance indicated that the differences between the 2 methods were not affected by race (P = 0.70), age (P = 0.71), or sexual maturation (P
0.13).
|
As shown in Figure 2
, on average, DXA tended to overestimate %FM by 3.9%. Because the bias remained unchanged across the entire range of %FM measures, the overestimation could be nullified by subtracting 3.9% from all DXA estimates of %FM if agreement with the 4-compartment model is the desired outcome. With this adjustment, the 95% limits of agreement became 6.7%, which means that on an individual basis, %FM could be underestimated or overestimated by 6.7% when DXA is used to estimate FM in a young, multiethnic female population. The 95% limits of agreement of DXA are an improvement over those of the skinfold-thickness equation (10%; 23) or total-body electrical conductivity (8%; 24). More importantly, the bias and limits of agreement of DXA are not affected by body fatness or age, unlike the total-body electrical conductivity methodology (24).
A previous study showed that DXA is a precise method for body-composition measurement (25), and in fact, several authors have considered DXA to be a reference method for this purpose (2628). These enthusiastic reports, however, have been countered by other studies showing that body-composition measurements made by DXA, particularly those of body fat, can be significantly affected by bone maturation (29, 30), age (8), sex (9), tissue thickness (31, 32), skeletal content of FFM (10), choice of instrumentation (11, 3335), and choice of software (34, 35). On the basis of the data reported in the latter studies, several authors have already cautioned against the use of DXA as a "gold standard" for body-composition measurements (12, 36, 37).
The inaccuracy of the body-composition measurements made by DXA could be attributed to the assumption of a hydration constant of 73% for FFM (12). Although the analysis algorithm used by Hologic does not include a fixed hydration constant per se, it is indirectly assumed when the mass attenuation coefficient for the lean mass is calculated. The hydration constant of 73% for FFM might be appropriate for healthy adults but would not be appropriate for children or adolescents. Hydration of FFM has been shown to be substantially higher among children and to change with maturation (14, 38). However, subsequent reports (1, 39) have shown that under normal and even most clinical conditions, errors associated with fat estimation by DXA due to variation in soft tissue hydration are small and should not affect its accuracy.
It is possible that the propagation of the individual measurement error and technical error from each of the body density, TBW, and BMC measurements could affect our conclusions. However, in a chapter elegantly written by Siri (15), %FM could be estimated within ± 1.5% by using a 3-compartment model if TBW could be measured with an error of ± 1%. Because BMC accounts for a small fraction of FFM, the addition of BMC in the 4-compartment model would not alter the accuracy of the %FM estimation outlined by Siri. This conclusion is confirmed by a recent study (40) indicating that the additive errors in the multicompartment model did not offset the improved accuracy of fat estimations over that obtained by using densitometry or body water measurement alone. Because the measurement and technical errors associated with our TBW measurements based on H218O dilution are better than ± 1%, the propagated error of the %FM estimations of our subjects would be well within ± 1.5%. Therefore, the measurement and technical errors associated with the body density, TBW, and BMC measurements in our study would not alter our conclusions.
It is well known that the body composition of growing children is different from that of mature adults (41). The algorithms used by the Hologic DXA instrument are based on adult proportions. Thus, for the younger or smaller children and adolescents in this study, these algorithms may be less accurate. Furthermore, the calculation of %FM used by the Hologic DXA instrument is relative to the attenuation ratios or R ratios used to define the 100%FM and 0%FM values. These choices may be less accurate when applied to children and adolescents. Another possible source of error is that the bone detection algorithms used for adults are less accurate for children, thus resulting in the misclassification of bone into the soft tissue compartment, which then alters the true lean-to-fat ratio.
It has been documented that the density of FFM (23), the skeletal content of FFM (13, 42), and body fat distribution (4345), all of which affect the accuracy of body-composition measurements by DXA, are different between young African American and white females. Young African American females have a higher BMC (13) but lower amounts of total, visceral, and subcutaneous adipose tissue than do young white females (4345). Asian women also have more subcutaneous adipose tissue and a different fat distribution than do white women (46). These racial differences in body composition may further complicate the accuracy of the body-composition measurements made by DXA.
Our results show that the bias and limits of agreement of DXA are independent of age, ethnicity, or %FM. The limits of agreement of 6.7% for an individual young female with an average %FFM of 76.1% (the average %FFM4C of all young females in Table 1
) will translate into an overestimation or underestimation of FFM of 8%. However, the same limits of agreement for an individual young female with an average %FM of 23.9% (the average %FM4C of all young females in Table 1
) will be exaggerated into an overestimation or underestimation of body fatness of 28%. Thus, although DXA, with proper calibration, is an appropriate method for estimating body fatness in a group of young females, its accuracy for an individual in this group is suboptimal. The applicability of our conclusion to DXA instruments made by other manufacturers or to DXA instruments that use different scan modes and software is not known. For example, the Lunar DXA instrument has pediatric software and uses different scan modes in different weight brackets. However, the accuracy and precision of the Lunar DXA instrument for the estimation of %FM in a multiethnic group of young females are not known and warrant further investigation.
| ACKNOWLEDGMENTS |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
V. K Haas, J. R Allen, M. R Kohn, S. D Clarke, S. Zhang, J. N Briody, M. Gruca, S. Madden, M. J Muller, and K. J Gaskin Total body protein in healthy adolescent girls: validation of estimates derived from simpler measures with neutron activation analysis Am. J. Clinical Nutrition, January 1, 2007; 85(1): 66 - 72. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. S Rogers, A. R Ness, C. D Steer, J. C. Wells, P. M Emmett, J. R Reilly, J. Tobias, and G. D. Smith Associations of size at birth and dual-energy X-ray absorptiometry measures of lean and fat mass at 9 to 10 y of age. Am. J. Clinical Nutrition, October 1, 2006; 84(4): 739 - 747. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E Williams, J. C. Wells, C. M Wilson, D. Haroun, A. Lucas, and M. S Fewtrell Evaluation of Lunar Prodigy dual-energy X-ray absorptiometry for assessing body composition in healthy persons and patients by comparison with the criterion 4-component model Am. J. Clinical Nutrition, May 1, 2006; 83(5): 1047 - 1054. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. B. Sopher, J. C. Thornton, J. Wang, R. N. Pierson Jr, S. B. Heymsfield, and M. Horlick Measurement of Percentage of Body Fat in 411 Children and Adolescents: A Comparison of Dual-Energy X-Ray Absorptiometry With a Four-Compartment Model Pediatrics, May 1, 2004; 113(5): 1285 - 1290. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Gately, D. Radley, C. B. Cooke, S. Carroll, B. Oldroyd, J. G. Truscott, W. A. Coward, and A. Wright Comparison of body composition methods in overweight and obese children J Appl Physiol, November 1, 2003; 95(5): 2039 - 2046. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |