2 Supported by a grant from the National Health Service (West Midlands) Research and Development Scheme, United Kingdom, with additional funding from the Engineering and Physical Sciences Research Council; by the Grant Agency of the Czech Republic under project number 203/00/0632; by the Royal Society through a Joint Project Grant supporting the essential collaboration between DS and PS; and by the North Staffordshire Medical Institute (BE). 3 Address reprint requests to SJ Davies, Department of Nephrology, North Staffordshire Hospitals Trust, Princes Road, Hartshill, Stoke-on-Trent, ST4 7LN, United Kingdom. E-mail: simondavies1{at}compuserve.com.
| ABSTRACT |
|---|
|
|
|---|
Objectives: The objectives were to show the efficacy of the new FA-MS method in measuring TBW in healthy subjects and to compare these measurements with values derived from multifrequency bioelectrical impedance analysis, skinfold-thickness (SFT) measurements, and both recent and historical published regression equations.
Design: After baseline measurement of breath deuterium abundance, 24 healthy subjects ingested 0.3 g D2O/kg body wt. A second breath sample was taken after 3 h to measure the increase in deuterium, from which TBW was calculated. Bioelectrical impedance analysis was carried out with a multifrequency analyzer, and SFT was measured by a single trained observer. Methods were compared with the use of Pearsons correlation coefficient and Bland-Altman analyses.
Results: TBW measures obtained by all methods were highly correlated (r = 0.950.98, P < 0.001), especially those between FA-MS, SFT measurement, and recent regression equations. The mean values obtained were within 2% of those published for age-matched control subjects and varied by 16% when all methods were compared. Systematic bias was greatest when FA-MS was compared with bioelectrical impedance analysis, which tended to underestimate TBW in smaller, female subjects. No bias related to subject size was observed in a comparison of FA-MS with SFT measurement or with more recent regression equations.
Conclusions: FA-MS is a simple and effective new approach to TBW measurement in healthy subjects. The difficulty of using population-derived equations to estimate TBW in individual subjects is emphasized.
Key Words: Total body water deuterium isotope dilution flowing afterglow mass spectrometry bioelectrical impedance analysis breath test, noninvasive measurement
| INTRODUCTION |
|---|
|
|
|---|
Several approaches have been taken to the measurement of TBW; most of them provide little more than an estimate, often relying on regression equations derived from population data. These approaches include skinfold-thickness (SFT) measurements (5), bioelectrical impedance analysis (BIA) (6), differential X-ray absorption (7), computerized tomography scanning (8), and indirect measures of body density such as underwater weighing or air displacement (9). However, the most direct and accurate method involves isotope dilution. In this approach, a known dose of an isotopic variant of water is ingested, and, after the elapsing of enough time to allow equilibration of the deuterium within the body water, a sample of blood, urine, or saliva is taken for analysis by stable-isotope mass spectrometry (10). Isotope dilution using D2O and H218O as well as radioactive tritiated water has been used extensively (10, 11).
Currently, the use of isotope dilution to measure TBW is limited to the research environment. A number of factors prevented its becoming a routine clinical method, including the inconvenience of blood sampling, the delay (typically, several days) in the results of mass spectrometric blood analysis, and the cost. A breath test using H218O dilution and breath C18O2 analysis was developed (10), but the cost of the isotopic variant of water is prohibitive. BIA is more immediate, but it is time-consuming and its accuracy is compromised, because it relies on experimentally established algorithms (12, 13).
We have developed a new method for measuring, online and in real time, the fraction of deuterium in the water contained in single-breath exhalations (14, 15). We call this new method of stable isotope analysis flowing afterglowmass spectrometry (FA-MS) (14). We showed, by using standard dilute mixtures of D2O in water [which actually results in HDO/H2O mixtures (16)], that this method allows measurement of the deuterium-to-hydrogen (D:H) ratio in water vapor with >1% accuracy and precision in a few seconds during a typical breath exhalation (17).
The noninvasive nature of FA-MS, its simplicity (no sample manipulation is required), and its accuracy and immediacy make it attractive as a clinical tool. It is our objective to establish this method for TBW measurements in the clinical environment and to use it in conjunction with the BIA method, which gives additional valuable information on intra- and extracellular fluid compartmentalization but cannot provide accurate absolute measurements of TBW. As a step toward this goal, we carried out a comparative study of the TBW in 24 healthy volunteers with the use of parallel deuterium-dilution and BIA measurements of TBW.
| SUBJECTS AND METHODS |
|---|
|
|
|---|
|
Another exciting feature of the FA-MS method is its ability to explore the kinetics of water dispersal between the various body compartments. Typical time constants for dispersal (upper intestinal tract to blood; blood to muscle) are 1020 min. Thus, by an exploitation of the high sampling frequency of breath deuterium in the FA-MS method, which can be several times/min if necessary, the breath-time profile of deuterium can be well described (15). This also allows the equilibrated breath deuterium content to be well established (14).
Baseline breath deuterium values were obtained for all 24 subjects before D2O ingestion and for the control subject (
156 ppm, consistent with the values in local tap water). The subjects then ingested a quantity of 99.9%-pure D2O in
100 mL of tap water; the dose of D2O equaled
0.3 g·kg-1 body wt. The actual individual doses of D2O were accurately weighed, and the D2O was washed down with an additional 100 mL of tap water. Breath samples were taken
3 h after the D2O ingestion, after which time our previous studies have shown that a near steady state of breath deuterium content is reached, which implies even dispersal among the body water (15). The TBW is simply measured from the increase in the breath deuterium content in relation to the volume of D2O ingested. This is explained in more detail in our previous publications (14, 15). Finally, the TBW values were reduced by 4% to account for the H-D exchange that occurs between HDO and the carboxyl, hydroxyl, and amino groups in body proteins (10, 20).
Total body water measurement by bioelectrical impedance analysis
BIAs were made with a multifrequency analyzer (model 4200, Xitron Hydra; Xitron Technologies Inc, San Diego) (21). Subjects were asked to lie supine on a couch for 10 min. After the skin was cleansed to ensure good contact, and electrodes were attached to the right wrist and ankle in a standardized fashion: the voltage-detector electrodes were placed at the line of the joint and the current-injection electrodes were placed over the metacarpalphalangeal and metatarsal phalangeal joints, respectively. These electrodes were then coupled to the Xitron Hydra, which was operated from a laptop computer, to initiate and store the readings. The acquisition of data typically took 15 s.
The Xitron Hydra measures resistance and reactance, from which are calculated the reciprocal impedance at 50 programmed frequencies. These frequencies are spaced between 5 and 1 MHz for calculation of the combined extracellular and intracellular fluid volumes and thus of TBW. With the use of the Cole-Cole plot to determine the theoretical resistance at infinite frequency (12), the TBW is calculated from the formula
![]() | (1) |
where V is volume, KB is a factor relating the relative proportions of the leg, arm, torso, and height;
F is the resistivity of the fluid; L is the body height; and R is the resistance. This method of measuring TBW has been validated in young and middle-aged healthy subjects with the use of isotope-dilution methods, with a correlation ranging between 0.89 and 0.99 (22, 23).
Total body water measurements by skinfold-thickness measurement
Each subjects height was measured with the use of a wall-mounted stadiometer to the nearest 0.1 cm, and the weight was measured on a balance-beam scale to the nearest 100 g. A single trained person (BE) whose technique had been validated (CV using 4 sites within published range:
9%; 24) took SFT measurements at 4 sites (triceps, biceps, subscapular, and suprailiac) on the right side of the body with a SFT caliper (Holtain Ltd, Dyfed, United Kingdom) according to the methods described by Harrison et al (25). Three measurements were taken at each site in rotation, and the mean was used for subsequent analysis. Body density was calculated from the age- and sex- specific formulas of Durnin and Womersley (9), and the percentage total body fat was derived (26). TBW was derived from percentage body fat by calculating 0.73 x (weight - body fat) (20).
Statistical analyses
The precision of the FA-MS measurements was calculated with the use of the SE of triplicate breath analyses that took into account the Poisson distribution of the detected ion counts. Comparisons between the methods of measuring or estimating TBW were made by calculation of Pearsons correlation coefficient, and the slope of the intercepts of the regression lines was tested for significant differences from 1 and 0, respectively. Bias and systematic differences between methods were assessed by the method of Bland and Altman (27). Comparisons of the mean differences between methods of measuring TBW according to sex were analyzed with the use of analysis of variance to test for sex-by-method interactions.
| RESULTS |
|---|
|
|
|---|
1 min, and once the post-D2O ingestion breath sample had been obtained, the TBW was calculated immediately, because the FA-MS software has been developed to give instantaneous measurement of the deuterium abundance (P
pan
l, SCILIB library, version 4, 2001. Internet: http://scilib.webpark.cz). As would be expected, the baseline breath D:H essentially did not vary between subjects (±2 ppm), and it reflected the ratio in the local water supply (
156 ppm). One of the advantages of the FA-MS method is that, in contrast with conventional techniques, it does not require calibration (17), so in principle TBW can be measured to the required accuracy from a single post-D2O ingestion breath sample.
FA-MS measurements of TBW for each subject are shown in Table 1
, where they can be compared with estimates made with the use of BIA, SFT measurement, and the Watson (11) and Chumlea (18) formulas referred to previously. The TBW values from all 5 methods were highly correlated with each other (see Table 2 for summary), although the correlations were best in comparisons of FA-MS with the Chumlea formula and SFT measurement and least good in comparisons of BIA with the Chumlea formula and SFT measurement. The mean differences in TBW between methods are summarized in Table 3, and they vary from 1% to 6%. The best agreement was observed between FA-MS and the Chumlea formula, when both the mean differences and the spread of agreement (SD) are taken into account.
The effect of sex on the degree of the difference observed varied according to the method used. The largest discrepancies were observed in comparison between the Watson formula and other methods, whereas the least difference by sex was observed between FA-MS and SFT measurement.
Systematic bias according to subject size varied according to the methods compared (Table 4 and Figure 1). The bias was greatest when FA-MS was compared with BIA, such that the latter underestimated TBW in smaller, female subjects, whereas when the Watson formula was compared with the Chumlea formula, the former gave lower values for TBW in larger, male subjects than did the latter. When TBW, measured by either FA-MS or BIA, is related to BMI, clear separation by sex is evident (Figure 2), with different slopes of the regression lines. In other words, women generally have lower absolute TBW as well as a lower percentage body water relative to body mass than men have, as a result of womens higher body fat content. This suggests that the relatively poor agreement between these methods of TBW determination that is seen in smaller, female subjects is related to the different fat content in women, which is not adequately accounted for by the formulas used to derive TBW from BIA. The least systematic bias was seen between the SFT measurement, FA-MS, and Chumlea formula methods.
With these observations taken together, there appear to be 2 principal reasons for disagreement between these various methods, and both are related to sex differences. On the one hand, better agreement between FA-MS and the Chumlea formula appears to result from the closer values obtained in the men, because of the relatively higher absolute values for TBW obtained by the more recent regression equations. On the other hand, agreement is influenced by the degree of body fat, especially in women in whom BIA underestimated TBW. This latter problem is less evident with the use of SFT measurement, which by definition takes fat distribution into account.
| DISCUSSION |
|---|
|
|
|---|
The accuracy of the new FA-MS method was established previously (17) at close to 1%. The absolute measurements of TBW obtained in this study are in general agreement with those reported in the literature (11, 18). The mean TBW reported for women, after correction for deuterium exchange between HDO and body proteins, was 32.3 kg, representing 47.5% of body weight. This value is within 1.5 kg for age-matched women reported in series by Watson et al (11), Cohn et al (30), Piers et al (13), and Chumlea et al (31). As the data for these studies were collected over the last 25 y, it would seem that this aspect of body composition has been stable for white women of European origin, but such may not be the case for men. The mean value for TBW reported here for men was 46.2 kg (54.5% of body weight), which compares well to recent population studies of Chumlea et al (45.8 kg, 55.5%; 31) and Piers et al (45.3 kg, 58%; 13) but these data appear to be different from the older data reported by Watson et al (38.2 kg; 11) and Norris (40.8 kg) (as cited in Chumlea et al, 31) in comparing age-matched groups. When we compared the TBW measurements made with FA-MS with those derived from the regression equations published by Watson et al and Chumlea et al, we found that the former gave poor agreement in the men, whereas the latter showed good agreement for both sexes. The explanation for this observation would seem to be clear. More recent estimates of TBW in men of this age group (
age: 62 y) would indicate that TBW is relatively well preserved with aging, which perhaps reflects better nutrition than was seen in earlier generations. This possibility draws attention to another risk in using population-derived regression equations to measure TBW in individual patients, namely, that populations can change with time.
The secondary objective of this study was to compare the measurements of TBW made by the FA-MS method with estimates made by the BIA and SFT measurement methods. Despite the strong correlations found in comparisons of these methods (Table 2
), significant systematic differences were observed. In the comparison of FA-MS and BIA, these differences were most marked in smaller subjects, which suggests that the algorithms used to calculate TBW by the Xitron Hydra are overcompensating for the influence of body fat in patients who are less obese. Prediction of body fat from BIA methods can vary considerably according to the regression equations used (32), with some studies showing a range of overprediction or underprediction by an average of 6% in either direction in the same women (33). Among the factors confounding the BIA approach is the effect of regional variation in fat distribution. There is a tendency for fat to be overestimated in subjects with predominantly upper-body distribution and underestimated in those with predominantly lower-body distribution (33). The systematic differences between FA-MS and SFT measurement were not influenced by sex and body size. There was, however, a consistent difference between the 2 methods, such that FA-MS gave, on average, a value for TBW that was 2 kg higher than SFT measurement gave. The reason for this difference is not certain, but it cannot be ascribed to sex differences. The accuracy of SFT measurement is subject to several potential variances, including the accuracy of measurements, which can vary by 9% even in experienced hands (24), the applicability of the population-derived equations for determining body density in individual persons, and the assumption that the hydration of fat-free mass is 0.73 in all persons. That figure, originally determined from cadaveric measurements, is subject to variability (4%) (20).
|
The intention now is to produce a small FA-MS instrument (Trans Spectra Limited, Stoke-on-Trent, United Kingdom) for use at the laboratory bench and convenient siting in the clinical environment (eg, in renal dialysis units) to perform D:H measurements in the water contained in single-breath exhalations. It is anticipated that any person, after slight training, could operate such an instrument and thus obtain an accurate and precise measurement of TBW. It would then be sensible to perform TBW measurements by using FA-MS in conjunction with BIA measurements. An accurate BIA algorithm could be derived for each person, and this would avoid the problem of applying population-derived formulas that clearly result in demonstrable inaccuracies in TBW determination.
|
|
|
|