ORIGINAL RESEARCH COMMUNICATION |
1 From the Departments of Dietetics (PLMC-R), Clinical Chemistry (BVK), and Surgery (PBS), University Hospital Maastricht, Maastricht, Netherlands.
2 Supported by the University Hospital Maastricht. 3 Reprints not available. Address correspondence to PLM Cox-Reijven, Department of Dietetics, University Hospital Maastricht, P Debyelaan 25, PO Box 5800, 6202 AZ Maastricht, Netherlands. E-mail: ncox{at}pmdi.azm.nl.
| ABSTRACT |
|---|
|
|
|---|
Objective: The primary aim of the study was to assess the validity of BIS in patients with gastrointestinal disease. The secondary aim was to compare different screening methods for identifying nutritionally depleted patients.
Design: In 70 patients with gastrointestinal disease, BIS measurements were performed and fluid compartments were calculated from extrapolated resistance (R) values of extracellular (RECW) and intracellular (RICW) water based on mixture equations. Results were compared with total body water and ECW, which were measured by dilution methods. Patients were classified as depleted if their actual fat-free mass (FFM) was <90% of their ideal FFM (iFFM). Different screening approaches for depletion were compared: the impedance vector method and the resistance index (RI) in relation to iFFM (RI/iFFM).
Results: Total body water and ICW were largely underestimated (4 L) in the not-depleted patients but not in the depleted patients. In both groups, ECW was slightly overestimated (0.60.7 L). The critical frequency was 60% higher and the membrane capacitance was 40% lower in the depleted than in the not-depleted patients. The impedance vector method failed to identify depleted patients, whereas the proposed ratio at R50 (R measured at 50 kHz) and Rinf (R at infinite frequency) resulted in comparable sensitivity (86%) and specificity (7380%).
Conclusions: BIS measures of body fluids are influenced by the presence of depletion, probably because of alterations in the electric properties of the body at the cellular level. However, for screening purposes, the use of the ratio (L2/R)/iFFM at R50 and Rinf may be of value.
Key Words: Patients gastrointestinal disease depletion body composition bioelectrical impedance spectroscopy BIS total body water extracellular water intracellular water screening
| INTRODUCTION |
|---|
|
|
|---|
Bioelectrical impedance analysis (BIA), which involves measurement of the impedance of the human body to an alternating current, may be of value. However, many different approaches are currently in use. Single-frequency BIA (SF-BIA) measures body impedance at one frequency, usually 50 kHz, from which total body water (TBW) is calculated on the basis of regression equations that incorporate the resistance index (RI = L2/R), where L is body length and R is resistance. These equations were found to be population specific, which complicates the clinical use of SF-BIA (24). Furthermore, it has been concluded that in cases of fluid imbalance, which is often present in severely ill patients, SF-BIA is not valid (5). Bioelectrical impedance spectroscopy (BIS) may overcome the limitations of SF-BIA, because it measures the fluid compartments ECW and ICW separately. BIS measures the impedance at a range of frequencies from which the resistance of the extracellular (RECW) and intracellular (RICW) fluid compartments are extrapolated. These resistance values are used to calculate ECW and ICW separately. This approach also differs from SF-BIA in that mixture equations based on physical models are often used instead of regression equations (6). In view of the differences between the 2 approaches, BIS seems more appropriate for clinical applications than is SF-BIA.
For screening purposes, L2/Rwithout the use of population-specific prediction equationsmay be of value. Because the size of fluid compartments is proportional to L2/R, comparison of the RI at different frequencies with normal values of fluid compartments may allow the identification of nutritionally depleted patients. The applicability of this ratio as a screening tool has not been studied.
The BIA vector method described by Piccoli et al (7) is another approach that does not require prediction equations. This method uses the plot of the impedance vector components, R and reactance measured at 50 kHz. Studies have shown that this method is useful in subjects with a high body mass due to fat or fluid overload (8) and in dialysis patients undergoing fluid removal (9). However, it is not clear whether the method is clinically useful in discriminating between depleted and nondepleted patients, because no studies of this approach in patients with gastrointestinal diseases or in depleted patients have been published.
The primary aim of the present study was to assess the validity of BIS for measuring fluid volumes in patients with gastrointestinal disease by comparing BIS with dilution methods and to test the validity of BIS in the presence of weight loss and nutritional depletion. The second aim was to assess the applicability of the RI (L2/R) and the impedance vector method as screening instruments for nutritional depletion by comparing the sensitivity and specificity of these approaches with results from BIS measurements and computation of the volume of body compartments.
| SUBJECTS AND METHODS |
|---|
|
|
|---|
Protocol
All measurements were performed in the morning on the first day of admission. Patients had fasted overnight. After the patients voided, bioelectrical impedance measurements and then anthropometric measurements were performed. Body weight was measured to the nearest 0.1 kg, and length was measured to the nearest 0.1 cm. Wrist circumference was measured to the nearest 0.1 cm with a metal tape. Blood and saliva samples were taken and then the patients drank an indicator dilution mixture of deuterium oxide and sodium bromide. Four hours after ingestion of the indicators, blood and saliva samples were taken again. During the 4-h interval, patients were not allowed to eat, and drinking was limited to 200 mL and allowed only during the first 2 h.
Blood hematocrit and hemoglobin concentrations were analyzed according to routine hospital procedures. Patients were interviewed regarding their usual weight and weight changes during the previous year. On the basis of the degree of weight loss, patients were divided into 3 groups: no, moderate, and severe weight loss. On the basis of fat-free mass (FFM), measured by deuterium dilution in relation to ideal FFM, patients were divided into depleted and not-depleted groups.
Dilution methods
TBW was measured by deuterium oxide dilution (TBWdil), and ECW was measured by sodium bromide dilution (ECWdil). In the morning, postabsorptive patients drank a mixture of 20 g deuterium-labeled water (99.84 atom percent excess) and 30 mL of 150 mmol NaBr/L. Before and 4 h after ingestion of the indicators, saliva and blood samples were obtained for deuterium and bromide analyses, respectively. Deuterium was analyzed by a mass spectrometric method described by van Kreel et al (10). Bromide was analyzed by ion chromatography as described by Wong et al (11). Plasma was not deproteinized by the use of a filter, but by acetonitrile.
TBWdil was calculated by correcting the deuterium dilution space with 4% for exchange of labile hydrogen. ECWdil was calculated by correcting the bromide dilution space with 0.90 for nonextracellular distribution of bromide and with 0.95 for the Donnan equilibrium. The accuracy of TBW measurements by deuterium dilution is reported to be 1.5% (12, 13). The bromide dilution method for ECW is estimated to have an accuracy of 5% (13, 14). Intracellular water (ICWdil) was calculated as TBWdil - ECWdil. From TBWdil, the FFM of patients was calculated by assuming a hydration factor of 73.2:
![]() | (1) |
Bioelectrical impedance spectroscopy
BIS measurements were performed just before the subjects drank the indicator mixture. The subjects lied on a bed with their legs separated and arms abducted from the body. A tetrapolar electrode (3M red Dot Ag/AgCl; 3M Health Care, Borken, Germany) arrangement, as described by Lukaski et al (15), was used. Impedance was measured with a Xitron 4000B bioelectrical impedance spectrometer (Xitron Technologies, San Diego) with the use of 48 frequencies ranging from 5 to 500 kHz. Measurements were made at the right and left sides of the body. Mean resistance values were used for analysis of the data. From all 48 measured resistance and reactance values, RECW and RICW were extrapolated by using the Xitron curve-fit software. All fits were classified as "good" or "excellent." RECW and RICW were used in Xitron equations based on the Hanai mixture theory for calculating body fluid compartments (6, 16). In healthy persons, TBW and ECW calculated by this method (TBWBIS and ECWBIS, respectively) agreed well with values from the dilution methods (for TBW: r = 0.95, SEE = 1.33 L; for ECW: r = 91, SEE = 0.90 L) (6).
The resistance at infinite frequency (Rinf) was calculated as follows:
![]() | (2) |
The curve-fit software also provides values for critical frequency (Fc) and membrane capacitance (Cm). Fc is the frequency at which the reactive component of the impedance reaches maximum values. Cm is a measure of the ability of membranes to store electric charge. Cm of the whole body is determined by total cell surface area and membrane thickness and porosity. Furthermore, Cm is also affected by the aspect ratio (length to cross-sectional area) of the bodys conductor (6). To study the effect of depletion on membrane properties, we calculated the Cm index, which removes the effect of the aspect ratio (by inclusion of length2) and of the total cell surface area (by inclusion of FFM measured by deuterium dilution):
![]() | (3) |
Calculations of ideal body weight and ideal fat-free mass
The patients wrist circumference was used to calculate frame size. On the basis of body length and frame size, the ideal body weight was calculated according to tables published by the Metropolitan Life Insurance Company (17). From the ideal body weight, ideal FFM was calculated in kg with the use of 80% of ideal body weight in men and 70% in women as reference values. If the FFM of patients, measured by the dilution method, was smaller than 90% of their ideal FFM, the patients were classified as depleted.
Weight-loss categories
On the basis of their usual weight and reported weight loss preceding the study, patients were divided into 3 categories. Patients with weight loss
5% over 1 mo or 10% over 6 mo were classified as having severe weight loss. Patients with no weight loss were classified as having no weight loss, and all other patients were classified as having moderate weight loss.
Screening methods
Three different screening methods were compared: screening based on BIS with mixture equations, screening based on the BIA vector method, and screening based on the RI.
Screening based on BIS with mixture equations
Patients were classified as depleted by BIS if FFMBIS was <90% of the ideal FFM.
Screening based on the BIA vector method
The resistance and reactance values of patients measured at 50 kHz were recorded and normalized for height. The sex-specific bivariate normal distribution of resistance/height and reactance/height published by De Palo et al (18) and by Piccoli et al (19) were used to assess the presence of depletion. For interpretation of the patients vectors normalized for height, the graphic method was used. The sex-specific nomogram was divided into 4 sectors. Patients with vectors in the upper or lower right 95% tolerance ellipse or outside the nomogram at the right side were classified as depleted according to the instructions in the manual of the soft tissue analyzer of Akern Bioresearch (Florence, Italy).
Screening based on the resistance index
The RI at different frequencies was calculated as L2/R, in which R represents the resistance measured at 50 kHz (R50), or RICW or Rinf. To study the applicability of the RI as a screening tool for nutritional depletion, its value in relation to ideal FFM was assessed by the ratio RI/ideal FFM in both the depleted and not-depleted patients. The mean values and 95% CIs (x ± 2 SE) of this ratio were calculated in both groups. If the CIs did not overlap, the mean distance between the upper limit in the depleted group and the lower limit of the nondepleted group was set as the threshold value. Screening was performed by comparing the value of RI/ideal FFM with the threshold value. Patients were classified as not depleted or depleted if this value was higher or lower than the threshold value, respectively.
Data analysis
Fluid volumes measured by BIS were compared with values measured with the dilution methods by Bland-Altman analysis (20). The correlations between methods for TBW and ECW were calculated and a plot of the difference between methods and their mean value was made. A paired Students t test was used to calculate whether the difference between methods was significant.
To study whether BIS prediction errors (dilution - BIS) were associated with the physical characteristics of the subjects, Pearsons correlation coefficients (r) between the prediction error and age, length, weight, BMI, ECW/TBW, body fluid volumes, and FFMdil were calculated.
The effect of weight loss and presence of depletion on body composition, BIS variables, and BIS prediction errors was tested by analysis of variance. Differences between weight-loss categories were tested by independent-samples Student t tests, with Bonferroni-corrected P values. To study the effect of depletion on these variables, differences between the depleted and not-depleted groups were tested by independent-samples Student t tests.
To compare the ability of the different screening methods to identify patients with nutritional depletion, the sensitivity (percentage of depleted patients classified as depleted) and specificity (percentage of not-depleted patients classified as not depleted) of the different approaches were calculated with the dilution method as the gold standard. For all tests, P values < 0.05 were considered significant.
| RESULTS |
|---|
|
|
|---|
|
|
|
|
|
Effect of depletion
Of the 70 patients, 14 were classified as depleted on the basis of their FFM in relation to their ideal FFM. This group had significantly lower values for body weight, BMI, TBW, ECW, and ICW and had a higher ECW/TBW value as measured by the dilution methods (Table 1
). Blood hemoglobin concentrations and hematocrit were lower in depleted patients. TBWBIS, ECWBIS, and ICWBIS were also significantly lower in the depleted group than in the not-depleted group. Fc was 60 kHz higher in the depleted patients, and the Cm was 0.59 nF lower. The Cm index was also significantly lower in the depleted group. The BIS prediction errors were different in the 2 groups. Compared with the not-depleted group, in which BIS underestimated TBW and ICW and overestimated ECW, BIS prediction errors were significantly smaller for TBW and ICW in the depleted group.
Comparison of screening methods
Individual RI/ideal FFM values for Rinf, R50, and RICW in the depleted and not-depleted groups are shown in Figure 4
. The CIs for RICW in the 2 groups overlapped; therefore, (L2/RICW)/ideal FFM was not used for further analysis. The ratios at 50 kHz and at infinite frequency were significantly lower in the depleted than in the not-depleted group, and the threshold values were 1.02 and 1.29, respectively (Figure 4
). Application of these values for screening of the total population resulted in sensitivity values comparable with results obtained from screening based on FFM values measured by BIS with use of mixture equations (Table 3
). The specificity found for screening based on R50 values was higher than for BIS or for screening based on Rinf.
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Validity of bioelectrical impedance spectroscopy
Results of the present study show that BIS with the use of mixture equations underestimated TBW and ICW and only slightly overestimated ECW. Although the errors in TBWBIS were not associated with the degree of weight loss, they were smaller in depleted patients than in not-depleted patients; this effect was not found for ECW. Overall, ECW predictions were accurate with an acceptable SEE of 2 L, which is comparable with results found by others (6, 2123). The prediction of TBW was less accurate with an SEE of 4 L. The findings of other studies confirm our finding that BIS predicts ECW better than does TBW (2428). Our finding that TBW and ICW were largely underestimated in the not-depleted patients but not in the depleted patients limits the clinical applicability of BIS. It can be argued that in the patients classified as depleted, the severity of illness was greater than in the not-depleted patients and that the severity of illness, not depletion, is responsible for these findings.
The degree of weight loss is often used as an indicator of depletion. In the present study, however, body weight but not TBW was significantly lower in the patients with severe weight loss than in the patients with no weight loss, which indicated that the weight loss was mainly the result of fat losses. The classification of patients as depleted or not depleted was based on the amount of FFM in relation to the ideal FFM. The 2 approaches obviously led to different classifications. In fact, 15% of the patients with no weight loss, 18% of the patients with moderate weight loss, and 24% of the patients with severe loss were classified as depleted. Weight loss was calculated from usual body weight, as reported by the patients, which may have been subject to considerable error because of the uncertainty of the patients about their usual body weight and because of the improper calibration of scales. Furthermore, weight loss may be masked by fluid accumulation. Therefore, weight loss alone is not a valid indicator of depletion in patients with gastrointestinal disease.
Depletion can be defined as a loss of BCM, the metabolically active part of the body, which leads to impairment of physiologic function. In this study we defined depletion as a condition in which patients have an FFM that is < 90% of their ideal FFM. We are aware that this is not the most optimal approach, but within the frame of practical alternatives it is most likely the most reasonable one. BCM cannot be measured directly; ICW can be used as a measure of BCM, but it has to be calculated as the difference between TBW and ECW. Because of error propagation, ICW calculations are by definition less accurate than are TBW or ECW measurements. Furthermore, no normal values for ICW specified for frame size are available at the moment. FFM is strongly related to BCM in cases of normal hydration. Patients in the present study did not show clear physical signs of fluid imbalance, although there was a small but significant increase in the ECW/TBW value in depleted patients. This means that FFM calculated from TBWdil contained relatively more ECW in the depleted than in the not-depleted patients. Consequently, the BCM of these patients may be more depleted than the FFM values indicated.
The finding that the BIS prediction errors for ECW and TBW are correlated with the absolute as well as with the relative size (expressed as ECW/TBW) of the fluid compartments has already been described in patients with altered fluid distribution (29). In acute diseases, clinical depletion is associated with ECW expansion in combination with ICW losses (25, 30). Because BIS prediction errors for both ECW and TBW were correlated with the true ECW/TBW value in the present study, but in opposite directions, BIS was unable to detect an increase in this ratio. This finding seriously limits the clinical use of BIS.
Dehydration as well as overhydration may affect conductive properties of ECW and ICW, resulting in changes in the specific resistance of these fluid compartments (31, 32). Changes in the composition of body fluids but not in their volume may also affect specific resistance. In the Hanai-based mixture equations, these values are set as constants based on measurements in healthy humans (6). For the clinical application of BIS, additional information regarding electric properties, electrolyte concentrations, and osmolarity appear to be necessary to adjust the used constants.
The findings that TBW was less accurately measured by BIS than was ECW and that TBW and ICW values, in contrast with ECW, were influenced by the presence of depletion indicate that factors related to measurements at high frequencies (used for measurement of TBW and ICW) are partly responsible for measurement errors. In several patient groups, depression of the phase angle at 50 kHz has been described (3335). It has been hypothesized that alterations in the electric properties of membranes are responsible. Impairment of membrane potential during illness may cause a decrease in Cm. In critically ill children, the reactance component of the impedance measured at different frequencies was significantly decreased (36). In our study, Cm calculated by the Xitron curve fitting software was lower on average than the values measured in healthy persons (1.25 compared with 2.18 nF) and Fc values were higher (112 compared with 60 kHz) (6). Furthermore, Cm was lower in the depleted patients than in the not-depleted patients and was lower in patients with severe weight loss than in those without weight loss. In the same groups, Fc was higher by 60%. Because Cm is also a function of total cell surface area, which is lower in depleted patients because of the loss of BCM, we calculated the Cm index. According to de Lorenzo et al (6), this index is a measure of cell porosity and thickness. Our results show that the Cm index is lower in patients with severe weight loss and in depleted patients. At present, it is not clear how the values for these parameters should be interpreted physiologically, but they indicate changes at the level of the membranes of cells during illness. Changes in the composition of ICW can also alter the dielectric properties of the capacitors formed by cell membranes and thus affect the Cm. These changes have no effects on resistances measured at low frequencies; however, these changes appear to affect resistances measured at frequencies above the Fc. It is likely that such changes are important in explaining the BIS prediction errors for TBW and ICW. Studying alterations in electric properties of cells during illness is necessary for a complete evaluation of the clinical applicability of BIS.
Screening methods
It was impossible to accurately identify depleted patients with the vector method described by Piccoli. Promising results with this method have been described for screening fluid overload or dehydration (8, 37).The low sensitivity of the method in our study may be caused by the fact that only height was used to normalize the measured R50. The use of the RI at 50 kHz (L2/R) divided by the ideal FFM appeared to be a screening instrument with high sensitivity and specificity. The method is refined by assessing the ideal FFM based on ideal body weight. The advantage of this screening method is that no population-specific prediction equations or complicated mixture equations are needed. Problems related to measurements in the high-frequency range, as discussed above, may be responsible for the fact that R50 values gave better results than did Rinf or the mixture-derived FFM values. It can be argued that a frequency of 50 kHz is too low for full penetration of the current through cells. This was especially true in the depleted patients in the present study in whom the Fc was 159 kHz. In fact, at this frequency, it is probable that only ECW plays a part in current conduction. Because of the strong interrelations between fluid compartments and the absence of severe fluid imbalance, this ECW measure is strongly related to the amount of FFM and explains the good results.
In conclusion, BIS with the use of mixture equations measures ECW accurately in patients with gastrointestinal disease, but the accuracy of ICW and TBW measurements is lower and influenced by the presence of nutritional depletion. Alterations in the electric properties of the body as a result of changes at the cellular level may play a role. A better understanding of the underlying physical phenomena seems necessary before widespread clinical use can be advocated. The impedance vector approach is not sensitive enough to screen patients for clinical depletion. Screening based on the RI at 50 kHz or based on Rinf in relation to ideal FFM has a high sensitivity and specificity and may be of value for clinical use in patients with normal fluid distribution.
| ACKNOWLEDGMENTS |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. S. Powers, L. Choi, R. Bitting, N. Gupta, and M. Buchowski Rapid Measurement of Total Body Water to Facilitate Clinical Decision Making in Hospitalized Elderly Patients J Gerontol A Biol Sci Med Sci, June 1, 2009; 64A(6): 664 - 669. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Earthman, D. Traughber, J. Dobratz, and W. Howell Bioimpedance Spectroscopy for Clinical Assessment of Fluid Distribution and Body Cell Mass Nutr Clin Pract, August 1, 2007; 22(4): 389 - 405. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Sun, C. R French, G. R Martin, B. Younghusband, R. C Green, Y.-g. Xie, M. Mathews, J. R Barron, D. G Fitzpatrick, W. Gulliver, et al. Comparison of multifrequency bioelectrical impedance analysis with dual-energy X-ray absorptiometry for assessment of percentage body fat in a large, healthy population Am. J. Clinical Nutrition, January 1, 2005; 81(1): 74 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Gupta, C. A Lammersfeld, J. L Burrows, S. L Dahlk, P. G Vashi, J. F Grutsch, S. Hoffman, and C. G Lis Bioelectrical impedance phase angle in clinical practice: implications for prognosis in advanced colorectal cancer Am. J. Clinical Nutrition, December 1, 2004; 80(6): 1634 - 1638. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Buchholz, C. Bartok, and D. A. Schoeller The Validity of Bioelectrical Impedance Models in Clinical Populations Nutr Clin Pract, October 1, 2004; 19(5): 433 - 446. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |